Material Stainless steel, Material No. 1.4301
Basket dimensions height approx. 430 mm
width approx. 500 mm
depth approx. 600 mm
Volume approx. 130 l
Max. weight of specimen approx. 20 kg
1.2.2 External Housing
Dimensions height 1 approx. 1990 mm
width approx. 1820 mm
width with door
hinge approx. 1850 mm
depth without
door approx. 900 mm
depth with removable
door approx. 1215 mm
Required clearances for
test cabinet height approx. 2010 mm
width approx. 1870 mm
depth ² approx. 1220 mm
Exterior finish two-coloured
grey-white RAL 9002
blue violet RAL 5000
Door manual sliding door
Door window approx. 450 x 500 mm
1 The height of the unit in operation with extended gear rack is approx. 2550 mm.
² If the observation door is removed, the required depth clearance is approx. 950 mm.
1.3 Installation Drawing
1.4 Equipment
Shelf 1 pc., made from stainless steel, adjustable in specimen basket
Port approx. 32 mm , port combined with the gear rack, moving with the basket
Refrigeration unit hermetic, water-cooled cascade
12 colour touch panel arranged ergonomically
Control Digital Measuring and Control System S!MPAC*
Interfaces 1 Ethernet interface (100/10 Megabit) for integration into network
1 USB interface for direct documentation of measuring data via USB stick (USB stick is not included) or as printer interface
High-/low temperature safety device as per EN 60 519-2 (1993), with separate sensor, thermal safety class 2, 1 in the hot chamber and 1 in the cold chamber.
1.5 Performance Data
Changeover time approx. 10 sec.
Temperature change of 10 kg ICs 15 min.
Temperature ranges
- Hot chamber
- Cold chamber
+60 °C to +220 °C
-80 °C to -10 °C
Temperature constancy, in time ±1 K
Calibration values -40 °C cold chamber and
+125 °C hot chamber
Note: